Skip to main content

HITSD

Adjustment of laser lines without opening the laser cover

Low tilt sensitivity

Lasers certified for an MRI environment till 3T

As indicated in its name, the HIT Single Diode includes only one diode per optical head, and it can be either red or green

The HITSD consists of one sagittal line and two cross lasers

(Possibility to add more lines (for example put a sagittal cross line)

What advantages does the HIT Single Diode offer compared to other laser systems on the market ?

– Adjustment of laser lines on all the three planes of movement is possible without opening the laser cover. This allows a better stability of the adjustment. Indeed, if the user needs to remove the cover to adjust the lasers, putting the cover back inflicts a constraint and unsettles them. With the CYRPA HITSD, only a simple wrench is necessary to adjust the laser, from the outside, without removing the laser cover.

– The tilt sensitivity is a key point when adjusting a laser manually. The less sensitive, the better. CYRPA HITSD has a very low tilt sensitivity which allows a more simple and more stable adjustment.

Manually fixed lasers are useful in every type of MRI room with compatible supports supplied by the room.

Request a Demo

Benefits & features

Red Laser

  • Type : Diode
  • Wavelength : 635nm
  • Maximum output power <1mW
  • Laser Class : Class II
  • Fineness of the line : < 1mm
  • Line Length : > 3m

Green Laser

  • Type : Diode
  • Wavelength : 532nm
  • Maximum output power <1mW
  • Laser Class : Class II
  • Fineness of the line : < 1mm
  • Line Length : > 3m

Manual Box

  • Overall Dimensions (Height x Width x Depth) : 220 x205 x 110 mm
  • Weight: 2 kgs
  • Power Supply: Internal 110/230 VAC.
  • Consumption of less than 500mA
  • Temperature: 15-30 °C
  • Type of control: Manual
  • Adjustable without removing the cover

Want to learnmore about HITSD?

Support services available

Our knowledgeable and experienced C-RAD Services team is here for you during the entire lifetime of the systems. We guide you from installation and clinical implementation, through preventive maintenance, support and continuous software upgrades.

Learn More

Related resources from learning center

Expanding Horizons of SGRT Explored by a C-RAD Expert

| Articles | No Comments
Daniel Speir, Chief Radiation Therapist, Willis-Knighton Health System, Shreveport, La. A newly published article by Jennifer Cook, C-RAD Clinical Applications Specialist at Vertec Scientific, a distributor of C-RAD products in…

C-RAD Opens New Experience Center in India

| Articles, Motion Management, Patient positioning, SGRT | No Comments
C-RAD is excited to announce the official opening of the C-RAD India Experience Centre, the first of its kind in Asia and the Middle East.   Opened in collaboration with…

C-RAD & Accuray: Partners for Progress in Breast Cancer Treatment

| Articles, DIBH, Motion Management, Patient positioning, SGRT, Workflow automation | No Comments
C-RAD is proud to be partnering with Accuray on a new VitalHold™ Breast Package for their Radixact System, integrating and automating a Deep Inspiration Breath Hold (DIBH) solution using Surface…
Close Menu